DESCRIPTION
BTI-DCM-100 conductivity meter measures the amount of electrical current or conductance in a solution. Conductivity is useful in determining the overall health of a natural water body. It is also a way to measure changes in wastewater procedures at water treatment plants.
•Rugged conductivity/TDS meter for efficient field water testing.
•Intuitive user interface for simple operation, reliable and accurate results.
•Trust your measurements - smart probes store all calibrations in the probe.
•Designed for demanding field operating conditions and ease of use.
•Complete convenient kit.
|
SAL FACTOR 0.75 (BY DEFAULT) & ADJUSTABLE MANUALLY |
|
|
CELL CONST ADJUSTMENT |
DIGITAL 0.05 TO 2.000 BY DEFAULT 1.000 |
|
TEMP COMPENSATION |
0 TO 70 DEGREE AUTO WITH TEMP PROBE?MANUAL WITH KEYS |
|
TEMP COEFFICIENT ADJ. |
0.00% TO 4.00% ADJUSTABLE BY DEFAULT 2.00% |
|
CALIBRATION |
MANUAL |
|
REFERANCE TEMP |
20 DEGREE C & 25 DEGR C (SELECTABLE) |
|
MEASURING FREQUENCY |
100 HZ & 1KHZ AUTO SELECTION TEMPERATURE |
|
RANGE |
-20.0 DEGREE C TO +200 DEGREE C |
|
RESOLUTION |
0.1 DEGREE C |
|
ACCURACY |
+/-% OF RANGE +/- 1 DIGIT |
|
CALIBRATION |
TWO POINT MANUAL |
|
SENSORS |
PT 100 PROB |
|
DATA ENTRY |
BY SIX SOFT TOUCH KEYS WITH AUDIBLE SOUND |
|
STORAGE |
1000 SAMPLES |
|
REAL TIME DISPLAY |
24 HOUR MODE WITH DATA |
|
INTERFACE |
RS-232 CENTRONICS & COMPUTER |
|
PRINT FORMAT |
ALL DATA DATE WISE |
|
DIGITAL OUTPUT |
SERIAL RS-232 PARALLEL CENTRONICS |