The MODEL BIO GENE BTI-GCN-5890 is a versatile gas chromatograph designed to be compatible with a wide range of detectors. It supports both packed column and split/splitless capillary injectors, offering flexibility for diverse analytical applications.
This system is fully compatible with the Agilent 6890N detector, along with related detector controllers, chromatographic workstation software, and automatic sampling systems, ensuring seamless integration into existing laboratory setups.
Engineered for precision and durability, the BTI-GCN-5890 delivers excellent analytical performance, stable operation, and long-term reliability. Its user-friendly design and robust construction make it suitable for routine as well as advanced laboratory use.
MODEL BIO GENE BTI-GCN-5890 is widely used for the analysis of oil extraction and refining, petrochemical, fine chemical, food safety, environmental protection, electronic high-purity gas industry, scientific research institutes, QC, safety supervision, public security forensic and national security and so on.
1. The temperature control accuracy can be precise to 0.01 ℃ by the new integrated digital electronic circuit with high control precision and reliable performance.
2. It has the functions of self-diagnosis for boot, stopwatch (for determining the flow), timer operation, storage protection for power off, keyboard locking, anti-interference, network data communication and remote control.
3. It can gets lower limit of detection by unique injector system design.
4. You can choose automatic/manual gas six-way valve injector, headspace injector, thermal desorption injector, cracking furnace injector, methane conversion furnace and auto-injector.
5. The program heating and the furnace temperature precision can be controlled stable and fast.
6. Optional detectors are FID, TCD, ECD, FPD, NPD and PDHID. All detectors can control the temperature individually. FID is easy to assemble and disassemble for cleaning or changing the nozzle. TCD stable time is ≤20min.
| Column oven | |
| Furnace dimensions | 28 x 30 x 18cm |
| Temperature operating range | Higher than The room temperature of 5℃-450℃ |
| Temperature setting | 0.1℃ programmed heating rate 0.1℃ |
| Maximum heating speed | 120℃/min |
| Temperature stability | It will be 0.01℃, when the environment temperature changes 1℃. |
| Programmed temperature steps | 7 steps |
| Column loss can be compensated (dual channel column) | |
| Injectors | |
| Various injector available: packed column injectors, split/splitless capillary column injectors or cool on-column injectors | |
| Detectors | |
| FID | |
| Maximum using temperature | 450℃ |
| LOD | ≤5 x 10 -12 g/s [n-C16] |
| Linear dynamic range | ≥107 |
| Drift | ≤5 x 10 -13A/30min |
| Noise | ≤2 x 10 -13 A |
| TCD | |
| Maximum using temperature | 400℃ |
| Sensitivity | ≥10000mv.ml/mg [n-C16] |
| Linear dynamic range | 104 |
| Drift | ≤ 10 0uV/30min |
| Noise | ≤20uA |
| FPD | |
| Maximum using temperature | 250℃ |
| LOD | ≤2 x 10 -13 g/s(P),≤5 x 10 -11 g/s (S) |
| Linear dynamic range | ≥103(P),≥102(S) |
| Noise | ≤2 x 10 -12 A |
| Drift | ≤4 x 10 -11A/30min |
| ECD | |
| Max using temperature | 350℃ |
| Drift | ≤5 0uV/30min |
| LOD | ≤1 x 10 -14 g/ml(y-666) |
| Linear dynamic range | ≥104 |
| Noise | ≤20uA |
| NPD | |
| Max using temperature | 400℃ |
| LOD | ≤5 x 10 -13 g/s(P),≤1 x 10 -12g/s (N) |
| Noise | ≤4 x 10 -13 A |
| Drift | ≤2.5 x 10 -12A/30min |
| Linear dynamic range | ≥103(P),≥103(N) |
| PDHID | |
| LOD | ≤5ppb |
Auxiliary facilities for option
DK-4001: a new generation of automatic headspace injector.