Bio Gene Conditioning Chambers are designed for precise control of temperature and humidity conditions across a wide range of industrial and laboratory testing applications. Manufactured by Biotechnologies Inc., these chambers are ideal for evaluating product performance under controlled environmental variables such as heat, cold, and moisture. Engineered to meet international standards, BioGene Conditioning Chambers offer programmable cycling, fast ramp rates, and uniform air distribution - making them suitable for aerospace, electronics, automotive, and materials testing environments. Rugged, accurate, and customizable, BioGene delivers dependable conditioning chamber solutions trusted by R&D and quality assurance labs worldwide.
ELECTRICAL PANEL
STANDARD PROTECTIONS
SOFTWARE CAPABILITIES:
TEMPERATURE SENSER:
Bio Gene Conditioning Chamber with HMI is designed with chamber feeding through Electric panel which is separated from cooling equipment to prevent the water condensation on the electrical components.
AIR-CONDITIONING CHANNEL
Evaporators, drainage, and the dry to prevent.
The Conditioning will be Done By Air Circulation To Maintain Uniform Temperature By Means Of low Noise Fan By Externally Mounted Motors.
ADJUST FEET AND WHEEL
Adjust feet: 4 castors.
CONTROL PANEL
Bio Gene Conditioning Chamber with HMI is designed with automated operation with provision for manual over ride. The operation will be via touch screen using appropriate software (electronic software control compatible with Windows OS
RECORDER & PC CONNECTIVITY (OPTIONAL)
TEMPERATURE LIMITERS
INNER CHAMBER
Bio Gene Controlled conditioning system Interior construction made of 18 SWG stainless steel 304 grade.
Internal construction material industry hinges and sensor mounts not corrode and
No joints and sharp edges inside workplace.
All the joints will be leak proof for prevention of air and water 'leakage.
Exterior body
Exterior body made of18 SWG Mild steel sheet with electrostatic powder coated finish.
Built in control panel provided on exterior body without any projection and All the hinges and door locks made of stainless steel.
CNC based made outer body With fine finishing joints without any projections of fixtures.
Appropriate airflow for homogeneous sample temperature.
Lift air tightness using suitable silicon seals.
Refrigerant: CFC free environmental friendly refrigerant.
Appropriate Compressor mounding to prevent noise and vibration.
All wiring to be done with suitable identification.
Electric panel separated from cooling equipment to prevent the water condensation on the electrical components.
Chamber has Internal Air Circulation, And All The Heaters And Cooling Coil is Located Behind/Below The False Sheet Of The Chamber And Outside The Working Space.
Test Space is Welded Vapor Tight With Rounded Edges And With Suitable Shelf Supports.
DOOR
Bio Gene Conditioning Chamber with HMI always provides door locking system having both hinged and sliding door as per requirement. And for safety purpose and experimenting purpose, The door is firmly locked by lever, thermal sealing to be ensured and Door opening and closing is manual and auto mode provision is optional
Multi Pane Glass Window.
Bio Gene Controlled conditioning system provided with Double wall insulated comprising all the sub-systems for operation and Robust type chrome finished heavy duty SS hinges fitted on the right hand side of the chamber with front opening single door.
INSULATION : Glass wool 150 mm thick. All seals on the controlled conditioning system and door for leak proof will be of silicon rubber double gasket for hermetic sealing better quality rubber.
Bio Gene Controlled conditioning system provided with Full front opening double walled insulated double doors with glass window of 250 x 250mm and silicone rubber gasket -50 to +100°C temperature envelope as well as humidity envelope of 10-98% RH.
The door hinges and toggle type locking arrangement. Door micro-switch off all running functions in case of door opening.
Lockable Front Opening Left hinged, With Single Door Latch Easy to Open.
Viewing Glass is Properly Sealed To Avoid leakage And provided With Heating Provision To Prevent Condensation. Windows size (250x250 mm),
Bio Gene Controlled conditioning system Heating, refrigeration and humidity system controlled conditioning system provided with flame proof fittings.
Viewing window : 2 nos. viewing window of 250 mm x 250 mm with multi-panned safety glass. One on each door.
Defogger's will be provided with auto cut off. Arrangement of light in the controlled conditioning system with control switch on the panel for easy viewing through window.
Controlled conditioning
system flooring & drain : will be strong enough to withstand static pressure of650 kg/m2. Shall has necessary arrangements to drain out condensed water, out of the controlled conditioning system.
TRAY
Bio Gene Conditioning Chamber with HMI platform trays have a provision to slide inside outside for the placement of the samples.
ILLUMINATION
Internal Light With External Control for monitoring purpose
KEY PARAMETERS
TEST ENVIRONMENTAL CONDITIONS
TEMPERATURE RANGE
Cabinet 1(Cold cabinet) specification: Lowest temperature with Refrigeration system :(-)-50 to 100ºC.
Bio Gene Controlled conditioning system provided with electrical equipment mounted on the controlled conditioning system such as air circulation motor, inspection lamp, door limit switch, thermocouples etc shall be flameproof.
Heaters will be Fong cold portion and terminated in flameproof junction box.
Make |
Bio Gene |
Model |
BTI-TSC-45 |
Operating temperature range |
-50°C to +100°C |
Controlled conditioning system temperature range |
-50°C (±1°C) to +100°C (±1°C) |
accuracy |
+1°C |
Shelves l weight |
175 Kg each. |
Stored holding |
36 Nos |
internal dimensions |
2000 mm (W) x 1500 mm (D) x 1500 mm (H) |
SAMPLE |
|
Diameter |
200 mm |
Length |
up to 1000 mm |
dimensional tolerances to limit within |
±5%) |
accommodate |
up to 36 numbers |
Cooling rate: |
: 2°C - 10°C/Minute averaged from 100°C |
Heating rate: |
2°C - 10°C/Minute averaged from 0°C to 100°C |
Relative Humidity |
10% to 98% + 3% |
RH in climatic range |
I-25°C to +55°C |
Low humidity |
<3% f:3% RH |
Power supply |
3 phase, neutral earth 415 volts + 5%, 50 Hz AC |